Safety & Handling
Linkam stages have EXTREMELY fragile
platinum leads; please do not touch or move.
When heating above 300
o
C, stages require
water cooling. Contact DP staff (2-9490).
Regarding gas purging:
Do not use hydrogen or helium (thermal
conductivity)
Dilute any reactive gases with 95% inert gas
Use only mixtures containing less than 20%
oxygen
Flow less than 60 CC/min
If using LN2 sample cooling, please follow all
APS procedures for safe handling of LN2.
The black capillary tube on the LN2 dewar lid
is fragile; please handle carefully.
Disconnect the LN2 dewar from the stage
before heating above 300
o
C.
Hardware Operation
Make all connections
If cooling sample with LN2:
Fill the dewar approximately 2/3 full.
The LNP95 must be switched on before the
T95/CI95 system controller.
The stage chamber needs to be purged of
air before starting a cooling experiment.
(over)
Quick Guide - 1
TS1500
DSC with cooling
Water cooling
THMS600
Please review the complete User Guide to ensure safe and effective operation.