Safety and Handling
The heating element may be fragile. Exercise care when loading and
unloading samples.
Use as little sample as possible to reduce thermal load and avoid damaging
the heating element.
Heating over 300
o
C
Stages require water cooling when heated above 300
o
C. Contact DP staff (2-9490 or
dp@aps.anl.gov) to obtain water circulator.
Disconnect the LN2 dewar from the stage before heating above 300
o
C. Leaving the tubing
connected can damage the tubing and/or heater.
Regarding gas purging:
Do not use hydrogen or helium (due to high thermal conductivity)
Do not use a gas mixture containing more than 20% oxygen
Dilute any reactive gases with 95% inert gas such as nitrogen or argon
Use a gas flow of less than 60 CC/min to avoid overloading the heating element
Failure to observe these guidelines may result in heating element failure.
If you plan to use the stage in a manner not recommended above, please purchase your own
heating element.
If using LN2 sample cooling, please follow all APS procedures for safe
handling of LN2.