Sample Cooling with LNP95
If using LN2 cooling, please follow all
APS procedures for safe handling of LN2.
The black capillary tube on the dewar lid
is fragile. Please handle carefully.
Fill the dewar approximately 2/3 full.
The LNP95 must be switched on before the T95/CI95 system
controller.
The stage chamber needs to be purged of air before starting a
cooling experiment.
See purging procedure on next page
Disconnect the LN2 dewar from the stage before heating above
300
o
C. Leaving the tubing connected can damage the tube and/or
heater.