5: Warning Messages
5.1 Fatal Error Messages
If a FATAL message is shown, the 500 automatically powers down
and cannot be used until the appropriate remedial action has been
taken. The most likely occurring examples of such a message are:
Instrument Too Hot Instrument Too Cold Change/Charge Battery
The Error Code X messages are explained in the following table. If
any of these occurs, please contact service personnel for attention.
lion alcolmeter
2 Fuel cell amplifier signal is too high
3 Fuel cell output too low during instrument calibration
4 Gas flow rate applied during breath flow sensor calibration was too low
5 Charge pump did not generate the required voltage
6 Temperature below -6°C
7 Calibration or checking attempted below 10°C
8 Calibration or checking attempted above 30°C
16 Fuel cell offset voltage has drifted too low
17 Blockage in sampling system
18 Used in Factory Mode only
19 Used in Factory Mode only
Error: Code 3