lion alcolmeter
®
500 Mk IV User Handbook
43
5. The instrument now calibrates itself, fully automatically:
…..
and then switches off. The process is now complete.
6. Before returning the 500 to service, you may wish to CHECK
its accuracy of calibration [Section 7 of this
User
Handbook], to verify that the above process has been
carried out satisfactorily. Make a record of the process.
11.5 The Calibration ADJUSTMENT Date
This is not normally stored or printed, but only that information
relating to the last Calibration CHECK.
Please Wait
Calibrating
Calibration
Adjusted
Please Wait