Guidance and Manufacturers Declaration – Electromagnetic
Immunity
Immunity
Test
IEC
60601
Test
Level
Compliance
Level
Electromagnetic
environment -
guidance
Conducted RF
IEC 61000-4-6
3Vrms
150 KHz
to 80
MHz
Not
Applicable
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the 500, including
cables, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
d = (3.5/V
1
)P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
d = (1.2)P 80 MHz to
800 MHz
d = (2.3)P 800 MHz to
2.5 GHz
where P is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and d is the
recommended