EasyManua.ls Logo

Loctite PULSE EQ RC50 - Page 26

Loctite PULSE EQ RC50
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
Enter Setting screen
- “Select 0 bar Cal. (2) for ST. Cal. (Set calibration), then check the actual value of
digital regulator (internal parts in the
reservoir, see the right figure).
- Compared the difference between the
actual value of digital regulator and 0 bar,
then click “+”/“-“(3) to increase/decrease
the value until the reading of “actual value
of digital regulator” is “0” bar. For larger
changes, touch the number field (3) and enter a new value directly.
Note that the 0 bar setting is the critical one, there is a range of about 0 370 of
the number, where the reading is close to 0(the prop reading always shows a
small increase, it is acceptable that there is some uncertainty of the display
reading of about +/- 0.02bar(0.0002Mpa) considering the accuracy of pressure
regulator.). The correct adjustment is achieved by increasing the number field (3)
until the prop regulator starts to increase from zero point.
- Click the (4) to record the actual reading as 1st “reference” point.
- “Select 5 bar Cal. (6) for ST. Cal., it will take several 10s for the pressure to
stabilize after setting to 5 bar. then check the actual value of digital regulator.

Related product manuals