EasyManua.ls Logo

LongerPump BT100-1F - User Manual

LongerPump BT100-1F
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
BT100-1F Longer Peristaltic Pump
Baoding Longer Precision Pump Co.,Ltd.
BaodingLongerPrecisionPumpCo.,Ltd
ADD: Building A,Chuangye Center Baoding
National High - Tech Industrial Development Zone
Baoding,Hebei,China 071051
TEL: 86 - 312 - 3110087 3138553
FAX: 86 - 312 - 3168553
E - mail: longer@longerpump.com
Www.longerpump.com
BT100-1F PERISTALTIC PUMP
OPERATING MANUAL
Question and Answer IconNeed help?

Do you have a question about the LongerPump BT100-1F and is the answer not in the manual?

Overview

The BT100-1F Longer Peristaltic Pump is a versatile and user-friendly device designed for precise fluid transfer and dispensing applications. It combines a 128x32 LCD display with a membrane keypad and a rotary coded switch for easy and prompt operation. The pump's speed can be adjusted manually or automatically through its external control interface.

Function Description:

The BT100-1F pump offers two primary working modes: Flowrate and Dispensing.

  • Flowrate Mode: In this mode, the pump delivers fluid at a set flow rate. The flow rate can be adjusted by turning the knob, and the current flow rate and speed are displayed on the LCD. The pump's running direction (clockwise or counter-clockwise) can also be changed.
  • Dispensing Mode: This mode allows for precise dispensing of specific volumes. Users can set the dispensing volume, the number of copies (total filling number), and the pause time between dispenses. The pump also features a "Back Suction" function to prevent liquid dripping after filling, which helps maintain accuracy.

The pump includes several control and setting features:

  • System Setting: This allows users to configure the pump head and tubing, enable or disable external control, and set the pump's RS485 address for multi-pump control.
  • Control Mode: Users can switch between Flowrate and Dispensing modes.
  • Dispensing Setting: Parameters such as dispensing volume, copy number, flow rate, and pause time can be set before a dispensing process.
  • Calibration: Both flow rate and dispensing volume can be calibrated to ensure high accuracy, especially when the pump's performance is outside the desired tolerance.
  • Prime Function: The pump can run at full speed for quick emptying, filling, and rinsing operations.
  • Memory Function: The pump automatically stores all running information.

Important Technical Specifications:

  • Acceptable Pump Heads: YZ1515X, YZ2515X, YZII15, YZII25, DG-1, DG-2, DG-4.
  • Flow Rates: Ranges from 0.16 µL/min to 500 mL/min.
  • Dispensing Volume: Ranges from 0.01 mL to 9.99 L.
  • Copy Number: 0 to 9999 (0 for continuous dispensing).
  • Pause Time: 0.1s to 99.9min.
  • Back Suction Time: 0 to 99.9s.
  • Display: 128x32 graphic LCD.
  • Operation Mode: Membrane keypad and Knob.
  • Direction Control: CW and CCW, reversible.
  • External Control Input: Supports Start/Stop, direction, and flow rate control (0-5V, 0-10V, 4-20mA, 0-10KHz optional).
  • External Control Output: Provides Start/Stop, Direction, and 0-10 kHz speed signals (OC Gate output).
  • Communication Interface: RS485 (up to 30 pumps can be controlled by a single computer).
  • Applicable Power: AC 90-264V, 50Hz/60Hz.
  • Power Consumption: <40W.
  • Operating Condition: Temperature 0 to 40 °C, Relative humidity < 80%.
  • Dimensions (L × W × H): 292.3 × 185 × 180.5 (mm).
  • Weight: 3.8 kg.
  • IP Rating: IP 31.

Usage Features:

  • Easy Operation: The combination of an LCD display, membrane keypad, and rotary coded switch ensures straightforward and quick operation.
  • Pump Head and Tubing Selection: Users can select suitable pump heads and tubing based on the desired flow rate and dispensing accuracy, with recommendations to keep the flow rate within 20% to 80% of the reference range for optimal performance. Thick wall tubing is preferred for longer service life.
  • Footswitch Function: An optional footswitch can be connected to control the pump's Start/Stop function, offering "Trigger" (press to start, press again to stop) or "Gated" (runs as long as pressed) modes. When a footswitch is connected, the Start/Stop key on the operating panel is disabled.
  • External Control: The pump can be integrated into a larger control system via RS485 serial communication or analog/pulse signal inputs, allowing for automated control of speed, direction, and Start/Stop functions.
  • Calibration for Accuracy: The ability to calibrate both flow rate and dispensing volume ensures high precision for various applications.

Maintenance Features:

  • Tubing Care: It is recommended to release the tubing from pressure when the pump is idle to prolong its service life.
  • Cleanliness: Keeping the rollers clean and dry helps extend the service life of both the tubing and the pump head.
  • Material Compatibility: The drive and pump head surfaces are not resistant to organic solvents and aggressive liquids, so caution should be exercised during use.
  • Grounding: Proper grounding of the pump is essential for operator protection, especially in moist environments.
  • Warranty and Support: The product comes with a one-year warranty, and support is available for troubleshooting and repairs.

LongerPump BT100-1F Specifications

General IconGeneral
BrandLongerPump
ModelBT100-1F
CategoryWater Pump
LanguageEnglish