Select pump head
and tubing refer to
thetableofpage4
External Control Input
Analog Signal Input function
Communication Control
The pump can connect to control computer (computer, PLC, SCM) through
RS485 serial communication module (shown as below). Please contact Longer
Company for communication protocol.
COM
B
A
11 12
Set the “External Control” in “on” state. External Control Module is shown as below.
Note:
Five kinds of standard external control module need to be ordered separately
according to special requirements.
5
4
3
2
1
External Control Module
Analog Control
4 - 20 mA Speed Control
0-5VSpeed Control
0 - 10 V Speed Control
0 10 KHz Speed Control-
External Control
DB15
Dispensing Mode operation procedure
Confirm the
volume to be
dispensed
Press to
return to running
interface
Exit Key
Operation Procedure
Select pump head and tubing refer to “Suitable Pump head, Tubing, Ref.
Flow Rate” table before working. Thick wall tubing is preferred for longer
service life.
Flow rate mode operation procedure
Confirm the flow
rate of the fluid
which need to be
pumped
Press to
enter
interface
to set pump head
and tubing
Knob
System
Setting
Enter
interface to
select
Control
Mode
Flow Mode
Press to
return to running
interface, turn
to select
needed flow rate
Exit Key
Knob
Press to
enter calibration
selection interface
and select
when
calibration is
needed
Knob
Flowrate
Set ,
press
to start the pump.
Input the actual
volume
Test Time
Start/Stop
Press to
return to running
interface
Exit Key
Select pump
head and tubing
refer to the table
of page 4
Press to
enter
interface
to set pump head
and tubing
Knob
System
Setting
Set
time
Back
Suction
Enter control
mode interface
to select
Dispensing
Enter
interface
to set all the
dispensing
parameters.
Dispense
Setting
Press to
enter calibration
selection
interface and
select
when calibration
is needed
Knob
Dispensing
Press
to start the pump.
Input the actual
volume
Start/Stop
Communication Control RS485
BT100-1F OPERATING MANUAL
BT100-1F OPERATING MANUAL