13 14
Pump is equipped with output port in order to monitor the status of pump. It
adopts DB9 male socket as output port (see as below). External control output
start/stop output, CW/CCCW output and pulse output ( 0 - 100 rpm
corresponding to 0 - 10 kHz). Output signal adopts optoelectronic isolation
circuit. A pull-up resistor and power supply must be added when using.
Terminals Definition
External Control Output
External Control Output Port Terminal Drawing: DB9
Footswitch Function
Connect to to control Start/Stop of
pump.
Footswitch External Control Input Port
Terminals Definition
0-5V,0-10V 4-20mA
Analog input to control the speed
,
CC/CCW Signal
Start/Stop Signal
1
4
3
4
2
0 5V, 0 10V, 4 20mA
External Control Input Module Connection Diagram
-- -
Pulse Signal Input Function:
0 - 10 kHz pulse input
+
-
4
5
CC/CCW Signal
Start/Stop Signal
+
-
3
1
+
-
2
COM1
COM
0 - 10 kHz Pulse Signal Input Connection Diagram
Analog input in 4 20 mA,0-5V,0 10Vexternalcontrolmodule
to control the speed of pump.
External control start/stop input. Open, the pump runs; Close, the
pump stops.
External control direction input. Open, the pump runs clockwise;
Close, the pumps runs counter clockwise.
COM of external control input.
--
Terminal 1
Terminal 2
Terminal 3
Terminal 4
COM of external control input
External control start/stop input. Open, the pump runs; Close, the
pump stops.
External control direction input. Open, the pump runs clockwise;
Close, the pumps runs counter clockwise.
Pulse input in 0 10 kHz pulse external control module to control
the speed of the pump.
COM of pulse input
-
Terminal 1
Terminal 2
Terminal 3
Terminal 4
Terminal 5
Terminal 1
Terminal 2
Terminal 8
Terminal 4, 6, 7
Start/Stop output. Pump outputs low level when it runs; pump
output high level when it stops.
CW/CCW output. Pump output low level when it runs clock
wise; pump output high level when it runs count clock wise.
Pulse output, 0 100 rpm corresponding to 0 10 kHz.
COM
--
Terminal 13
Terminal 5, 11
Note:
Footswitch is optional accessories. It can only control the start and stop of the
pump. The speed and direction of the pump are controlled by membrane keypad
and knob on the operation penal.
Start/Stop control. Close or input
pulse, the pump stops; Open or
input pulse again, the pump runs
COM when connecting terminal 5
and11together.
Connection Diagram for Footswitch
External Control Port
Relay
BT100-1F OPERATING MANUAL
BT100-1F OPERATING MANUAL