Chlorine L / M101
ENMethod Reference Book1.0 21
EN
Implementation of the provision free chlorine with liquid reagent
Select the method on the device
In addition, choose the test: free
10 ml
Sample
Fill 24 mm vial with 10 ml
sample.
Close vial(s). Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
Empty vial.
6 2
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add 6 drops DPD
1 Buffer Solution.
Add 2 drops DPD
1 Reagent Solution.