EasyManua.ls Logo

Lovibond MD 110 - KS4.3 T Method

Lovibond MD 110
369 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
K
S4.3
T
0.1 - 4 mmol/l K
S4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λ Measuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l K
S4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l K
S4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to K
S4.3
are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1
Revision status
KS4.3 T / 20
How to use
ENHandbook of Methods4
EN

Table of Contents