CyA T / M160
ENMethod Reference Book1.0 33
EN
Implementation of the provision Cyanuric Acid Test with Tablet
Select the method on the device
5 ml
5 ml
Fill 24 mm vial with 5 ml
deionised water .
Put 5 ml sample in the
vial.
Close vial(s).
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Press the ZERO button. Remove the vial from the
sample chamber.
Add CyA-Test tablet. Crush tablet(s) by rotating
slightly.
Close vial(s).