pH-value T / M330
ENMethod Reference Book1.0 41
EN
Implementation of the provision pH-value with Tablet
Select the method on the device
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s). Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Press the ZERO button. Remove the vial from the
sample chamber.
Add PHENOL RED
PHOTOMETER tablet.
Crush tablet(s) by rotating
slightly.
Close vial(s).