Faults
MXD70 Dissolved Oxygen
Setup and Operating Guide
- 57 -
Faults
E245 A Store A Checksum Error
E255 B
The Store A Save for the channel associated with this current output has become
corrupted. Check the current output's settings in the current output menu and
then save the settings again in the Channel's Store A in the Save/Restore menu.
E265 C
E275 D
E285 E
E295 F
E246 A Store B Checksum Error
E256 B
The Store B Save for the channel associated with this current output has become
corrupted. Check the current output's settings in the current output menu and
then save the settings again in the Channel's Store B in the Save/Restore menu.
E266 C
E276 D
E286 E
E296 F