Channel Setup
MXD70 Dissolved Oxygen
Setup and Operating Guide
- 17 -
Channel Setup
Temperature Input Sensor
Select the input channel’s temperature sensor type
for use with the sensor measurement’s automatic
temperature compensation.
If a temperature sensor is not connected to the input
channel then this menu item should be set to
disabled, else the temperature input error messages
will be shown.
Note. Even when disabled a manual temperature
compensation can be used.
/ – Select Option
EXIT
– Cancel
– Save Selection
Temperature Units
Sets the temperature units used.
/ – Select Option
EXIT
– Cancel
– Save Selection
Temperature Compensation Mode
To use temperature compensation from a measured
input select “Auto”, else select “Manual” to enable a
fixed value entry.
/ – Select Option
EXIT
– Cancel
– Save Selection