EasyManua.ls Logo

LTH Electronics MXD70 Series - Page 332

LTH Electronics MXD70 Series
631 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Channel Setup
- 16 -
MXD70 Dissolved Oxygen
Setup and Operating Guide
Channel Setup
Sensor Type
Sets the sensor’s input operating mode to either
Polargraphic or Galvanic.
Only available when probe type is set to “Custom”
/ – Select Option
EXIT
– Cancel
– Save Selection
Bias Voltage
For Polargraphic sensors, the polarising Bias Voltage
can be set using this menu.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value
Membrane Correction Factor
The membrane correction factor is specific to
each make of sensor and characterises the type
and thickness of the membrane material in
terms of how its permeability to Oxygen varies
with temperature.
The Correction Factor can be set at using this
menu.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value

Table of Contents

Other manuals for LTH Electronics MXD70 Series