Registers
Modbus RS485 Interface
Operating Guide
- 37 -
Registers
Conductivity Configuration Continued
4591 Get/Set Channel 1 Compensation
Slope*
FLOAT Temperature
Compensation Slope
Value
*(only available when
temperature
compensation set to in)
0 to 9.99%°C
5181 Channel 2
5781 Channel 3
4593 Get/Set Channel 1 Temperature
Mode*
INT Manual Temperature
Mode
*(only available when
temperature
compensation set to in
and temperature sensor
type not set to disabled)
1046 = Auto
5183 Channel 2 1047 = Manual
5783 Channel 3
4594 Get/Set Channel 1 Compensation
Input*
FLOAT Manual Temperature
Compensation Input
Value
*(only available when
temperature
compensation mode set
to manual)
-20.0 °C to 150.0 °C
5184 Channel 2 -4.0°F to 302.0°F
5784 Channel 3
4596 Get/Set Channel 1 Cable
Compensation
FLOAT Cable Length
Compensation Value
0 to 999 Meters
5186 Channel 2
5786 Channel 3
4598 Get/Set Channel 1 Input Filter INT Input Filter (Averaging) 1050 = Filter Out
5188 Channel 2 1051 = 10 Seconds
5788 Channel 3 1052 = 20 Seconds
1053 = 40 Seconds
1054 = 1 Minutes
1055 = 3 Minutes
1056 = 5 Minutes