Registers
Modbus RS485 Interface
Operating Guide
- 57 -
Registers
Electrodeless Conductivity Input Calibration
Electrodeless Conductivity Input Calibration
6650 Get/Set Channel 1 Cal Manual
Temperature
Input*
FLOAT Calibration Manual
Temperature Input
*(Only available when
Temperature
compensation Mode is
set to Manual)
-20.0°C to 150.0°C
7030 Channel 2 -4.0°F to 302.0°F
7410 Channel 3
6652 Get
Channel 1 Reading Slope
FLOAT Sensor Slope Value
80% to 120%
7032 Channel 2
7412 Channel 3
6654 Get
Channel 1 Temperature
Offset*
FLOAT Temperature Offset
Value
*(Not available when
Temperature Sensor is set
to Disabled)
25°C to +25°C
-13.0°F to 77.0°F
7034 Channel 2
7414 Channel 3
6656 Get/Set
Channel 1 Calibration
Access
INT Front screen Calibration
Access
1076 = Yes
7036 Channel 2 1077 = No
7416 Channel 3
6657 Get/Set
Channel 1 Calibration
reminder
INT Calibration Reminder 1076 = Yes
7037 Channel 2 1077 = No
7417 Channel 3
6658 Get/Set Channel 1 Calibration
Interval*
FLOAT Calibration Interval
Value
*(Only available when
Calibration Reminder set
to yes)
0 to 999 Days
7038 Channel 2
7418 Channel 3
6660 Get/Set
Channel 1 Calibration
Alarm Date*
INT Calibration Alarm Date
Value
*(Only available when
Calibration Reminder set
to yes)
1 to 31 Day
7040 Channel 2
7420 Channel 3
6661 Get/Set
Channel 1 Calibration
Alarm Month*
INT Calibration Alarm
Month Value
*(Only available when
Calibration Reminder set
to yes)
1 to 12 Month
7041 Channel 2
7421 Channel 3
6662 Get/Set Channel 1 Calibration
Alarm Year*
INT Calibration Alarm Year
Value
*(Only available when
Calibration Reminder set
to yes)
Max 2099 Year
7042 Channel 2
7422 Channel 3