LTSscale.com Page 14 of 28 REV F – 2018.04.09
Code 2100 - Keyboard Lockout
Note: Keyboard functions and the 90xx codes may be individually locked out.
Press keys 2100. Press ENTER key.
Top display line shows the item to be enabled or disabled. Bottom
display line shows the current status indicated by the asterisk.
Disabled or lockout = YES
Enabled or accessible = NO
To lock or unlock a function:
Use DEL key to toggle selection between YES and NO.
Press ENTER to accept selection and proceed to the next function.
Available Lockout Functions
Code 9083; Configure Remote Inputs
Code 9084; Configure Print
Code 9085; Set Time & Date – Option
Code 9085; Raw A/D, Battery Test
Menu ( lb./kg, Recall Tare )
Code 9087; Auto Shut Off and Backlight Timer
Code 9088; Piece Count/ APW
Note: in the event that the indicator locks up or does not respond properly, hold
down DEL and R keys.
Code 2104 - Expand Test Mode X10
Expand mode is used for testing with a finer graduation size than normally displayed.
Example: if the Graduation size is set for 5 lb. then expanded mode will display 0.5 lb. graduations.
To enter expand mode:
Press keys 2104. Press ENTER.
ZERO TRACK
Disables Zero Tracking in test mode. Re-enabled upon expand mode exit.
Press DEL key to select mode.
Press ENTER key to accept the selected setting.
Top display line shows the expanded weight.
Bottom display line shows the normal weight.
To exit and return to normal operation, simply press ENTER key.
Code 2105 - Restore Default Name
Restore the default name [ LTS SCALE ] to the top display line.
Press numeric keys 2105. Press ENTER key.
The top display line will display the default name in the Gross weigh mode.
LOCK KEYBOARD?
YES = LOCKOUT
!@#%^&*&^%$#@#@!
GROSS= 0 LB