LTSscale.com Page 21 of 28 REV F 2018.04.09
LTS DR2100A/AN Indicator
Code 9090 Single-Point Calibration
Single-Point Calibration can only be done after the Four-Point calibration has been completed at least
one time. If the display shows ERROR when this code is entered then the Four-Point calibration must
be completed first.
See Calibration section.
Code 9091 - View / Store Calibration Factors
View / Store Calibration Factors
This is not a substitute for test weight calibration.
Recorded Calibration factors can be entered to put the indicator back to a known
working condition. The Calibrations Factors stored will change after every Single-
Point and Four-Point Calibration. Recording of these factors is the last thing to be
done before putting the scale in service.
See Calibration section.
Code 9092 - View / Set Angle Zero and Span NTEP
See Out-of-Level Calibration.
View and set Angle Zero and Angle Span values.
Code 9093 View / Set Angle Factors
See Calibration section.
Code 9094 Out-of-Level Calibration
See Calibration section.