Model 26-1 Frisker User’s Manual Section 5
Ludlum Measurements, Inc. Page 5-7 April 2017
Secondary Units RATE/MAX Mode Alarm Point (Default 000) - Use
ON/ACK to select the value to adjust, and MODE to adjust the value. If the
Secondary Units is off, this parameter will be skipped. Otherwise, units will be
the same as selected earlier with Secondary Units. The ALARM LCD Segment
will be on to indicate an Alarm parameter. Set this Alarm Point to 000 to
disable. Available values are:
Ones Place (0-9)
Tens Place (0-9)
Hundreds Place (0-9)
Thousands Place (1 on or off)
Number of Decimal Places (0, 1, or 2– only available if k selected)
Range (k on or off – not available for mR/h or µSv/h)
Note: If the Secondary Units has changed to a value other than that used to
previously set this Alarm Point, the Alarm Point will be reset to 000.
Secondary Count Units (Default mR) - Use MODE to select. If the
Secondary Units is off, this parameter will be skipped. Otherwise, the available
values are dependent on the selected Secondary Units:
Secondary
Units
Secondary
Count
Units
Available
Secondary
Units
Secondary
Count
Units
Available
cps
cps
c
off
dpm
dpm
d
off
cpm
cpm
c
off
mR/h
mR/h
mR
off
Bq
Bq
d
off
µSv/h
µSv/h
µSv
off
Secondary Count Alarm Point (Default 000) - Use ON/ACK to select the
value to adjust, and MODE to adjust the value. If the Secondary Units is off,