GB Operating manual – multifunction measuring meter XA1000 / XP200 / XP400 • 08_2015 12
"Measuring mode – Automatic measurement" screen (integrated sensor)
42 mmHg: Displays the measured value according to the
specifications under 34 (e.g. in the measuring
unit millimetres of mercury) and under 39 (e.g.
as average value) (see fig. A., page 11).
43 Indicates the remaining recording duration for the automatic
measurement.
44 Opens a screen for selecting the recording duration for
measuring over a long period (see fig. E.).
45 Starts recording.
The key turns red once recording has started.
Note!
During an active recording it is not possible to switch off the
multifunction measuring meter with the "On/Off" key.
Automatic switch-off is also deactivated. First stop the
recording by pressing key 46 and then turn off the device.
46 Stops the current recording. The detected values are
automatically saved to the current measuring project.
47 Pauses the current recording.
Key 45 flashes.
Press this key again to continue recording.