31
N31U User manual
4004
4005
0...5
0...2
0
0
3 Reciprocal of the measuring quantity.
4 Square of the reciprocal of the measuring quantity.
5 Square root of the reciprocal of the measuring quantity.
Main displayed value on the main display eld
Value Description
0 Value of the measuring quantity, averaged and converted using the
mathematical functions and the individual characteristic
1 Minimum value of the measuring quantity, averaged and converted using
the mathematical functions and the individual characteristic, occurring
during the averaging window.
2 Maximum value of the measuring quantity, averaged and converted
using the mathematical functions and the individual characteristic,
occurring during the averaging window.
4006 0...6 2
Displayed resolution - position of the decimal point.
Value Format
0 000000
1 00000.0
2 0000.00
3 000.000
4 00.0000
5 0.00000
6 Automatic - the position of the decimal point is set for maximum
resolution.
4007 0...2 0
Contents of the bottom line of the display.
Value Description
0 Unit according to the selected unit (register 4008)
1 Measuring value without the mathematical operations.
2 Current time.
4008 0...56 0
Selection of the unit to be displayed in the lower line of the display on the main
screen (see register 4006). See point 5.1.
4009 0...9999 0
Access protection code to make changed in the conguration using the meter menu.
It would be required to provide a code each time when entering the meter menu in
case of entering a value higher than zero.