4-2 Calibration
1)Install the " Probe Plug " ( 4-1, Fig. 2 ) into the
" DO Socket " ( 3-13, Fig. 1 ).
2)Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
Press the " Function Button " ( 3-6, Fig. 1 ) once, to
let Display unit will show " %O2 " instead of " mg/L ".
3)Wait for approx. 5 minutes at least until the
display reading values become stable & no
fluctuation.
4)Use the two fingers to press the " REC Button " ( 3-4,
Fig 1 ) and HOLD Button " ( 3-3, Fig. 1 ) at the same
time, the display will show the following screen as
example, release the both fingers.
20.9
CAL
5)Press the " Enter Button " , the Display value will count
from 30 to 0, then return to normal measuring screen
and finish the calibration procedures. The complete
calibration procedures will take 30 seconds
approximately.
Press the " Function Button " ( 3-6, Fig. 1 ) once,
to let Display unit to return " mg/L ".
Calibration Consideration :
a.As the oxygen in air is 20.9 % typically, so use
the environment air 02 value for quick &
precise calibration.
b
.
Please process calibration procedures under
wide and ventilating environment for best effect.
10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lutron Electronics DO-5512SD and is the answer not in the manual?

Lutron Electronics DO-5512SD Specifications

General IconGeneral
BrandLutron Electronics
ModelDO-5512SD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals