User Interface Commands
4-6 FOT Lab Kit User’s Manual
AE Analog Output Activation
Enable all analog output channels which are attached to active
inputs. For each input channel that has been enabled using the PS
command, enable the corresponding analog output channel.
Information Request:
Syntax: <Esc>AE ? <CRLF>
Response: AE = ENABLED | DISABLED
Command Parameters:
Syntax: <Esc>AE = status <CRLF>
Range:
ENABLE Enable the analog outputs for all active inputs
DISABLE Disable all analog outputs
Examples: <Esc>AE = ENABLE <CRLF>
<Esc>AE = E <CRLF>
<Esc>AE = Disable <CRLF>
<Esc>AE = D <CRLF>