User Interface Commands
4-10 FOT Lab Kit User’s Manual
CC Calibration-Correction
The CC command is used to Enable or Disable application of the
Calibration-Correction Factors. Each installed temperature input
channel has associated calibration factors which are applied to the
measured temperature before it is reported.
These factors are set to typical values at the factory but must be
fine-tuned as needed to compensate for differences in probe type,
probe length and signal attenuation factors such as positioning and
optical fiber bend-radius inherent in your particular installation.
Please refer to Appendix A, “Calibration Procedure” for additional
details.
Information Request:
Command Syntax: <Esc>CC? <CRLF>
Response Syntax: CC = info
ENABLE Application of the Calibration-Correction Factors is enabled.
DISABLE Application of the Calibration-Correction Factors is disabled.
Parameter Modification:
Syntax: <Esc>CC=status <CRLF>
Range:
ENABLE (default) Enables application of Calibration-Correction Factors
DISABLE Disables application of Calibration-Correction Factors
Example: <Esc>CC = ENABLE <CRLF>
<Esc>CC = E <CRLF>