User Interface Commands
FOT Lab Kit User’s Manual 4-11
CT Calibration Temperature
The CT command is used to set the calibration reference
temperature used during the automatic calibration-correction
procedure.
Please refer to Appendix A, “Calibration Procedure” for additional
details.
Information Request:
Syntax: <Esc>CT ? <CRLF>
Response: CT = caltemp
Example: CT = 37.00 C
Parameter Modification:
Syntax: CT = caltemp <CRLF>
Range: caltemp New calibration reference temperature
The new reference temperature is specified in degrees and
hundredths of a degree within the working range of the device.
Temperature units are assumed to be the units previously set
using the UN command.
Value input is in "nn.nn" format. Do not include the units. If
the reference temperature entered is of an ambiguous format
e.g. "40" or "40.1" the system will correctly interpret the input
as "40.00" and "40.10" respectively. It is recommended,
however that values be entered in the standard format.
Example: <Esc>CT = 95.50 <CRLF>