EasyManua.ls Logo

Luxtron FOT Lab Kit - Page 41

Default Icon
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Interface Commands
4-12 FOT Lab Kit User’s Manual
DF Data Format
The DF command sets the serial-interface temperature output
format, either full or abbreviated.
The format selected depends upon the use to which the captured
data will be put. For visual clarity the abbreviated format is
recommended for most circumstances.
Information Request:
Syntax: <ESC>DF ? <CRLF>
Response: DF = format
Returns the current data reporting format.
Parameter Modification:
Syntax: DR = format <CRLF>
Range: FULL
FULL
Output is formatted for all channels without regard to the
number of active channels.
Examples:
All four input channels are active.
" 1: 20.42 C 2: 22.75 C 3: 21.25 C 4: 19.97 C"
Input channels 1 and 2 are active.
" 1: 20.42 C 2: 22.75 C 3: 4: "
Input channels 1 and 3 are active.
" 1: 20.42 C 2: 3: 22.75 C 4: "
ABBR
Output is compacted to display data for only active input
channels.
All four input channels are active.
" 1: 20.42 C 2: 22.75 C 3: 21.25 C 4: 19.97 C"
Input channels 1 and 2 are active.
" 1: 20.42 C 2: 22.75 C"
Input channels 1 and 3 are active.
" 1: 20.42 C 3: 21.25 C"
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com