User Interface Commands
4-22 FOT Lab Kit User’s Manual
SV Save
The SV command is used to copy the currently active user
parameter values and input-channel calibration factors to non-
volatile (NvRam) storage for use in subsequent sessions.
When user parameters are changed or when the instrument is
calibrated the resulting values are active only for the current session
i.e. until the instrument is reset using the CTRL+X command or
until the instrument is powered down. This allows you to
experiment with different parameter combinations without affecting
the stored parameters. If you wish to save the new calibration
factors and user parameters for subsequent sessions you must
explicitly save them using the SV command.
Information Request:
n/a
Parameter Modification:
Syntax: <Esc>SV <CRLF>
Response: none