User Interface Commands
FOT Lab Kit User’s Manual 4-21
ST Startup Mode
The ST command is used to select the "startup mode" indicating to
the device whether to power up into Standard mode or into
Standby mode.
In Standard mode the input circuits are enabled and the device
begins reporting temperatures as soon as data are ready.
In Standby mode the input circuits are idle i.e. samples are not
being taken and no data reports will be generated. In this mode the
user must explicitly enter Standard mode using the CTRL+R
command to enable data sampling and reporting.
NOTE Do not issue the CTRL+R command for at least 30 seconds after
power-up which will allow the device time to complete its startup
sequence.
See also the descriptions of the CTRL+T and CTRL+R commands.
Information Request:
Syntax: <ESC>ST ? <CRLF>
Response: ST = mode
Examples:
ST = ENABLE Temperature reporting starts automatically.
ST = DISABLE Reporting must be started manually.
ST = E Same as ST = Enable
Parameter Modification:
Syntax: <ESC>ST = mode
Range: ENABLE Begin reporting temperatures on power up.
DISABLE On startup wait for user to enable
temperature reports.