Operation Manual
touchTymp MI 24
and
MI 34
Version
In case of re-use of the single-use equipment you
enhance the risk of cross contamination!
MAICO strongly recommends to use Sanibel eartips only. In case you want to purchase
further disposables, please contact MAICO or your local distributor.
3.5 Components/Replacement Parts
Some reusable components are subject to wear with use over time. MAICO
recommends that you keep theses replacement parts available (as appropriate for your
touchTymp device configuration).
3.6 Troubleshooting
Table 2 Troubleshooting
Make sure the probe is connected to the back
of the device correctly and the brackets are
closed. Otherwise, follow the suggestions in
Probe tip.
1. Clean the probe tip as described in the
manual. If the system still does not run
proceed with step 2.
2. Use a new probe tip. If the system still
does not run proceed with step.
3. Change the complete probe and check if
the system is running.
Hold the Front key button for 10 seconds in
order to shut-off the device. Restart.
Turn off the device. Confirm/reconnect the
probe before restarting.
Transfer to
PC not
possible
Make sure the USB/PC-connection is
established, the PC module is opened and the
device and the connection icon is displayed
green. If the connection icon is not displayed
green, check if the right device is selected in the
Software Module (see Operation Manual of the
Software Module).
No license
Missing transducer
calibration
Combinations of
settings not allowed
Purchase license if wanted.
Calibrate transducer.
Verify settings are correct.
NOTE: If there are any problems that you cannot solve yourself, please, contact your
customer service. It will be helpful to use the function Export error log (see section 5.6.17)
to send the customer service the data needed for solving the problem.