6 Technical Data
This section offers you important information about
the touchTymp hardware specifications
connections
the pin assignment
immittance and audiometry calibration values
electromagnetic compatibility (EMC)
electrical safety, EMC and associated standards
6.1 touchTymp Hardware
The touchTymp is an active, diagnostic medical
product according to the class IIa of the EU medical
directive 93/42/EEC.
General Information About Specifications
The performance and specifications of the device can only be guaranteed if it is
subject to technical maintenance at least once per year.
MAICO Diagnostics puts diagrams and service manuals at the disposal of
authorized service companies.
STANDARDS
Medical CE-mark
Yes
Safety Standards
IEC 60601-1, ES 60601-1/A2/
CAN/CSA-C22.2 No 60601-1 Class I, Type B applied parts
EMC Standards
IEC 60601-1-2
Audiometer
Standards
Tone: IEC 60645-1 Type 3/ANSI S3.6 Type 3
Tympanometer
Standards
IEC 60645-5, Type 2/ANSI S3.39, Type 2
DEVICE SPECIFICATIONS
Power supply 24V
60W UE60-
240250SPAx
Consumption: 1.5A rms. at 90 VAC input and maximum load
Mains voltages and fuses: 100 VAC to 240 VAC ± 10 %
50 Hz to 60 Hz ±10 %
Mode of operation
Continuous