Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5 Page 22 June 23
6. Once the Bedpan / Urinal Bottle Washer Disinfector is “loaded”, break the infra-red beam again to close the door.
The display shows Shut
7. When the door is fully closed, 2 green LED’s will flash in the infra-red sensor area for 8 seconds. Break the beam again
to start a cycle.
The display shows rdY
If the 8 seconds has elapsed the flashing LED’s will stop and the Bedpan / Urinal Bottle Washer Disinfector will not start
for a cycle. The operator has to repeat the process from Step 5 (or press manual start).
Note: If this occurs the beam must be broken twice.
Note: On 10A models, if the steam generator temperature is not at 90°C after the beam has been broken or the manual
start button pressed, the display will show “CYCL”. The Green LED’s will stop flashing and stay on. When the steam
generator has reached temperature, the display will show rinS and the wash cycle will start automatically.