Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5 Page 5 June 23
1.1 Operating Cycle
Stage 1 Flushing / Cleaning
One rotating 180° back spray
Two fixed 60° top sprays
Two fixed urinal sprays
Rinse for 10 seconds
Detergent pump on for 7 seconds to add 30ml of concentrate to pump
Main pump on for 3 seconds to apply detergent
Pause for 40 seconds
Rinse for 10 seconds to wash detergent off
Stage 2 Thermal Disinfection
Steam is delivered into the sealed wash chamber
Steam continues to heat the surface temperature of utensils and chamber to a minimum of 90°C for 60 seconds
Stage 3 Cool Down Rinse
One 60° fine mist top spray delivers cold water direct from mains.
Utensils are cooled to 55°C for safe handling.
In accordance and complies with:
Cleaning Efficacy
AS/NZS 4187:2014, ISO 15883-1, ISO/TS 15883-5
Thermal Disinfection
AS/NZS 4187:2014, ISO 15883-1