Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5 Page 21 June 23
4.0 Cycle of Operation
1. Once the Bedpan / Urinal Bottle Washer Disinfector is installed into position with power and water connected, push
power on button.
2. When the Bedpan / Urinal Bottle Washer Disinfector is turned on via the display panel the display shows fill indicating
steam generator and cold-water tanks are filling to the high-water level.
At this stage the device will not commence a cycle until Step 5.
3. Once the steam tank water level probes sense the tank is full it will start heating.
4. The steam tank is heated to temperature 90°C - 92°C. The Bedpan / Urinal Bottle Washer Disinfector is now ready to
commence a cycle.
The display shows rdY (ready).
5. Break the infra-red beam by moving your hand between the infra-red indicators, this will open the door.
The display shows open.
Note: Whenever the door is opened the elements are in the off mode.