Hardware features
50
1. First, fill the cuvette with dispersant (6 ml minimum).
2. Insert the cuvette into the SV cell and then the instrument. Perform a background
and initial alignment.
3. The SV cell can then be removed and, using the syringe (with syringe tip attached),
extract an amount of dispersant.
4. Sample is now added into the cuvette, again to a 6ml minimum level.
5. Reinsert the SV cell into the instrument, perform another alignment and recheck
the obscuration.
6. If the obscuration is incorrect, remove the cuvette and add either more sample or
dispersant as required.
7. Repeat until the obscuration is correct, then continue with the measurement.

Other manuals for Malvern Mastersizer 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Malvern Mastersizer 3000 and is the answer not in the manual?

Malvern Mastersizer 3000 Specifications

General IconGeneral
Measurement principleLaser diffraction
Blue laser wavelength470 nm
Sample dispersion optionsWet and dry dispersion
Data acquisition rate10 kHz
Particle size range0.01 to 3500 microns
Laser typeHe-Ne and solid state
AccuracyBetter than 1%
RepeatabilityBetter than 0.5%
TechnologyMie and Fraunhofer theories
Sample dispersion unitsHydro MV, Hydro LV, Aero S
Optical modelMie theory
Power requirements100-240 V, 50/60 Hz
Detector arrangementLog-spaced array