Maintenanc e
97
Hydro SV procedures
Cuvette and stirrer bar cleaning general internal wash
Note: The cuvette must not be taken apart for cleaning. Users should only fol-
low the maintenance procedures specified.
Note: The cuvette and stirrer bar should be cleaned thoroughly before first use
and before any subsequent measurements are performed.
Stirrer bar initial cleaning (before first use)
The stirrer bar may still have some PTFE particles attached to it so will need to be cleaned
and any particles removed before being use.
1. Clean the stirrer bar with IPA or Acetone with a lint free cloth.
2. Leave the stirrer bar to dry on a clean sheet of lint free cloth.
Cuvette initial cleaning (before first use)
Use the wash station to clean the SV cuvette.

Other manuals for Malvern Mastersizer 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Malvern Mastersizer 3000 and is the answer not in the manual?

Malvern Mastersizer 3000 Specifications

General IconGeneral
Measurement principleLaser diffraction
Blue laser wavelength470 nm
Sample dispersion optionsWet and dry dispersion
Data acquisition rate10 kHz
Particle size range0.01 to 3500 microns
Laser typeHe-Ne and solid state
AccuracyBetter than 1%
RepeatabilityBetter than 0.5%
TechnologyMie and Fraunhofer theories
Sample dispersion unitsHydro MV, Hydro LV, Aero S
Optical modelMie theory
Power requirements100-240 V, 50/60 Hz
Detector arrangementLog-spaced array

Related product manuals