www.mc-techgroup.com PSS-5C | 1.02.02 41
21 Maintenance
Observe the facility- and process-specific safety measures before carrying out maintenance work!
WARNING
Dangerous voltage!
It is necessary to take the equipment off the mains before any
assembly, maintenance or repair work is carried out.
The frequency of the maintenance work depends on the operational process and can therefore only
be determined in each individual case.
All parts which might require maintenance work are easily accessible housed in the gas conditioning
system. The following parts might require maintenance work (see Figure 8):
Check the filter element of the fine filter FP-2T and replace if necessary.
NOTE
In order to protect downstream analyzers, the wet filter element
must always be replaced in the event of a condensate
breakthrough.
Check the tube of the SR25.2-W condensate pump every six months and replace if necessary;
NOTE
The complete instruction manual of the SR25.2-W is available at
www.mc-techgroup.com.
Check the diaphragm of the gas feed pump N3 KPE (optional N5 KPE or N9 KPE) every six months
and replace if necessary.;
NOTE
The complete instruction manual of N3 KPE, N5 KPE, N9 KPE is
available at www.mc-techgroup.com..
Check fan grids for contaminations and clean if necessary.