21
Range & Accuracy
Range Min Max Accuracy
(% of reading ±
counts)
V Read (low) 0.000 20.000 0.015% ±2
VDC Read (high) 0.0 250.0 0.05% ±2
VAC Read (high) 0.0 250.0 0.5% ±2
V Source 0.000 20.000 0.015% ±2
mV Read -10.00 75.00 0.015% ±2
mV Source -10.00 75.00 0.015% ±2
mA Read 0.000 24.000 0.015% ± 2
mA Source 0.000 24.000 0.015% ± 2
CPM Source 2.0 1000.0 0.05%
Hz Source 1 1000 0.05%
KHz Source 1.0 10.0 0.25%
CPM Read 0.0 1000.0 0.05% ± 1
Hz Read 0.0 1000.0 0.05% ± 1
KHz Read 0.00 10.00 0.05% ± 1
Range Min Max Accuracy
Ohms Read (low) 0.00 400.00 0.1 ohm
Ohms Read (high) 400.0 1500.0 0.5 ohm
1500.1 3200.0 1.0 ohm
Range Min Max Excitation Accuracy
Current
Ohms Source (low) 5.00 400.00 0.1 to 0.5 mA 0.15 ohm
5.00 400.00 0.5 to 3 mA 0.1 ohm
Ohms Source (high) 400.0 1500.0 0.05 to 0.8 mA 0.5 ohm
1500.0 3200.0 0.05 to 0.4 mA 1.0 ohm
Thermocouple Read and Source (All errors included)
Range Min Max Accuracy
J Thermocouple -200.0 0.0 0.6°C
0.0 1200.0 0.4°C
K Thermocouple -200.0 0.0 0.8°C
0.0 1370.0 0.5°C
T Thermocouple -200.0 0.0 0.8°C
0.0 400.0 0.5°C
E Thermocouple -200.0 0.0 0.5°C
0.0 950.0 0.4°C
R Thermocouple -20 0.0 2.49°C
0 500 1.7°C
500 1750 1.3°C
S Thermocouple -20 0 2.4°C
0 500 1.7°C
500 1750 1.4°C
B Thermocouple 600 800 2.1°C
800 1000 1.7°C
1000 1800 1.3°C