Thermocouple Read and Source (All errors included) (cont.)
Range Min Max Accuracy
L Thermocouple -200 0.0 0.45°C
0.0 900.0 0.4°C
U Thermocouple -200 0.0 0.7°C
0.0 400.0 0.45°C
N Thermocouple -200 0.0 1.1°C
0.0 1300.0 0.6°C
mV Read/Source -10.00 75.00 0.015% ±.02
mV
RTD Read and Source
Range Min Max Accuracy
Ni120 (672) -80.0 260.0 0.2°C
Pt100 (385) -200.0 800.0 0.33°C
Pt100(3926) -200.0 630.0 0.3°C
Pt100(3916) -200.0 630.0 0.3°C
Pt200(385) -200.0 630.0 0.8°C
Pt500(385) -200.0 500.0 0.3°C
500.0 630.0 0.4°C
Pt1000(385) -200.0 100.0 0.2°C
100.0 630.0 0.3°C
Cu10 -100.0 250.0 2.2°C
Cu50 -180.0 200.0 0.5°C
Cu100 -180.0 200.0 0.3°C
YSI400 15.00 50.00 0.05°C