EasyManua.ls Logo

Maxtec MaxO2 OM-25A - Page 4

Maxtec MaxO2 OM-25A
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
866.4.Maxtec www.maxtecinc.comII
NOTE: In order to obtain optimum performance from your MAXO
2
®
analyzer, all operation
and maintenance must be performed in accordance with this manual. Please read the
manual thoroughly before using the analyzer and do not attempt any repair or
procedure that is not described herein. Maxtec cannot warrant any damage resulting from
misuse, unauthorized repair or improper maintenance of the instrument.
WARNINGS
!
FAILURE TO COMPLY WITH THESE WARNINGS AND CAUTIONS COULD RESULT IN
INSTRUMENT DAMAGE AND POSSIBLY JEOPARDIZE THE WELL BEING OF THE PATIENT
AND/OR HEALTH CARE PROFESSIONAL.
»
The sensor of the MAXO
2
Oxygen Analyzer (MAXO2) has been tested with various
anesthesia gases including Nitrous oxide, Halothane, Isourane, Enurane, Sevourane
and Desurane and found to have acceptable low interferance, the device in its entirety
(including electronics) is not suitable for use in the presense of a ammable anesthetic
mixture. Only the threaded sensor face and sample inlet connection may be allowed to
contact such a gas mixture.
»
To avoid explosion, do not operate the oxygen analyzer in the presence of ammable
anesthetics or in an atmosphere of explosive gases. Operating the oxygen analyzer in
ammable or explosive atmospheres may result in re or explosion.
»
Never allow an excess length tubing near the patient’s head or neck, as such could result in
strangulation.
»
Before use, all individuals who will be using the MAXO
2
must become thoroughly familiar
with the information contained in this Operation Manual. Strict adherance to the operating
instructions is necessary for safe effective product performance. This product will perform
only as designed if installed and operated in accordance with the manufacturers operating
instructions.
»
Use only genuine Maxtec accessories and replacement parts. Failure to do so may
seriously impair the monitors performance. Repair or alteration of the MAXO
2
beyond
the scope of the maintenance instructions or by anyone other than an authorized Maxtec
service person could cause the product to fail to perform as designed.
»
Calibrate the MAXO
2
weekly when in operation and if enviromental conditions change
signicantly. (ie, Temperature, Humidity, Barometric Pressure. --- Refer to Calibration
section of this manual).
»
Use of the MAXO
2
near devices that generate electrical elds may cause erratic readings.
»
If the MAXO
2
is ever exposed to liquids (from spills or immersion) or to any other physical
abuse, turn the instrument OFF and then ON. This will allow the unit to go through its self
test and make sure everything is operating correctly.
»
Never autoclave, immerse or expose the MAXO
2
(including sensor) to high temperatures
(>70
°
C). Never expose the device to pressure, irradiation vacuum, steam, or chemicals.
»
To protect the unit from potential leaky battery damage always remove batteries when
the unit is going to be stored (not in use for 1 month) and replace dead batteries with
recognized name brand AA Alkaline batteries. Do not use rechargeable batteries.
»
This device does not contain automatic barometric pressure compensation.
»
Not for use in an MRI environment.

Related product manuals