9.0 SPECIFICATIONS
9.1 Base Unit Specifications
MeasurementRange: 0-100%
Resolution: 0.1%
AccuracyandLinearity: 1%offullscaleatconstanttemperature,R.H.and
pressure when calibrated at full scale.
TotalAccuracy: ±3%actualoxygenleveloverfulloperating
temperature range.
ResponseTime: 90%offinalvalueinapproximately15secondsat23°C.
Warm-up Time: none required
OperatingTemperature: 15°C-40°C(59°F-104°F)
Storage Temperature: -15°C - 50°C (5°F - 122°F)
Humidity: 0-95%(non-condensing)
Power Requirements: 2, AA Alkaline batteries (2 x 1.5 Volts)
Battery Life: approximately 5000 hours with continuous use
LowBatteryIndication: “BAT”icondisplayedonLCD
Sensor Type: Maxtec
®
MAX-250 galvanic fuel cell
ExpectedSensorLife: >900,000O
2
percent hours minimum 2-years in
typical applications
AModelDimensions: 3.0”(W)x4.0”(H)x1.5”(D)[76mmx102mmx
38mm]
A Weight: 0.4 lbs. (170g)
AEModelDimensions: 3.0”(W)x36.0”(H)x1.5”(D)[76mmx914mmx
38mm]Heightincludesexternalcablelength
(retracted).
AEWeight: 0.6lbs.(285g)
9.2 Sensor Specifications
Type: Galvanicfuelsensor(0-100%)
Life: 2-years in typical applications
13
W W W . M A X T E C I N C . C O M

Other manuals for Maxtec MAXO2+A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maxtec MAXO2+A and is the answer not in the manual?

Maxtec MAXO2+A Specifications

General IconGeneral
BrandMaxtec
ModelMAXO2+A
CategoryOxygen Equipment
LanguageEnglish

Related product manuals