This function allows the average load reading to be
displayed. The average starts being calculated when the
START threshold (% of full-scale) is reached and stops
being calculated when the load passes through the STOP
threshold.
To set average over time, press and hold the MENU key
until page 1 of the main menu appears. Using the UP and
DOWN keys, move the arrow cursor to AV/TIME and press
ENTER.
The display will show AV/TIME OFF and SET.
Press the ENTER key to change OFF to ON.
Press the DOWN key to move the arrow cursor to SET and
press the ENTER key.
The display will now show the START and STOP thresholds
and the values to which they are set (as a % of full-scale).
Any load reading above the START threshold will be
averaged over time. Averaging stops when the load reading
passes through the STOP threshold.
A diamond cursor will indicate which value is selected.
Use the UP and DOWN keys to change the value, press
and hold to scroll values. When the correct value is
reached press the ENTER key to set START.
Repeat procedure for setting STOP. The display will revert
back to AV/TIME sub-menu 1.
AV/TIME
The maximum duration of AV/TIME
calculation is approx. 22 minutes.
AV/TIME sub-menu 1
(SET)
AV/TIME sub-menu 2
(PERCENTAGE START/STOP)
1N.m
4N.m
3N.m
2N.m
0
Start Threshold
x Peak
TIME
S
top Threshold
Average