ActiveCare - Fixed
User Manual
MAN-ActiveCare REV02 ActiveCare Fixed USER MANUAL Page 22 of 25
Bed Transport
To transport an empty bed, lower the bed into the caster position and use the footboard or
headboard to move the bed. Do not move the bed using the rails, hand held pendant or assist bars
as handles. When bed is moved into intended space, raise bed off the casters.
NOTE: The bed is not intended to provide a means for transport of the occupant.
CAUTION: Never operate bed while in Transport Mode
Maintenance
Cleaning
The frame can be cleaned/disinfected with a disinfecting wipe. Disinfectant should always be
applied to a cloth; never spray directly on frame or controls. After applying cleaning agent, allow
the surface to dry prior to placing unit back in service. Clean frame as needed.
Maintenance and Service
NOTICE: RECOMMENDED TO PERFORM PREVENTIVE MAINTENANCE EVERY 12
MONTHS FOR BEST PERFORMANCE.
1. Inspect the frame for weld cracks or tears.
2. Check that all bolts, pins and fasteners are still in working order.
3. Perform functional test on the frame.
4. Inspect the casters for wear and tear.
Preventative Maintenance Inspections
Problem Resolution
Inspect bed frame for cracks, tears or bending,
particularly at welds and joints.
Remove bed from service. Contact Med-Mizer® customer
service.
Inspect electrical enclosures for cracks or other damage.
Remove bed from service. Contact Med-Mizer® customer
service.
Check that screws, bolts, pins and other fasteners are in
place, secured and properly tightened.
Remove bed from service. Tighten loose bolts and screws.
Contact Med-Mizer® customer service to order missing or
damaged fasteners.
Inspect actuators for excessive wear and misalignment.
Remove bed from service. Contact Med-Mizer® customer
service.
Inspect power cord for damage including the plug and
strain relief.
Remove bed from service. Contact Med-Mizer® customer
service to order replacement power cord.