What to do if Medela Medical Equipment temperature is too high?
- SShelley GarrettSep 8, 2025
If the Medela Medical Equipment displays 'Temperature too high', prepare a backup device.
What to do if Medela Medical Equipment temperature is too high?
If the Medela Medical Equipment displays 'Temperature too high', prepare a backup device.
What to do if Medela ThopEasy+ Medical Equipment canister is full?
If your Medela Medical Equipment displays 'Canister full', replace the canister.
What to do if Medela Medical Equipment filter is clogged?
If the Medela Medical Equipment displays 'Filter clogged', replace the canister and continue with 'on'.
What to do if Medela ThopEasy+ indicates leakage?
If the Medela Medical Equipment displays 'Leakage', check the patient status and the system for leaks, then continue with 'OK'.
What to do if Medela ThopEasy+ has leakage?
If you notice leakage in the Medela Medical Equipment, check the patient status and inspect the system for leaks. Then, continue with 'OK'.
How to unclog Medela ThopEasy+ Medical Equipment system?
If the Medela Medical Equipment system is clogged, check that the tubing is clear and not twisted. Also, replace the canister.
How to fix low battery on Medela ThopEasy+ Medical Equipment?
If the Medela Medical Equipment indicates a low battery, charge the battery by connecting the Thopaz device to the mains.
How to resolve an internal error on Medela Medical Equipment?
To address an internal error on your Medela Medical Equipment, switch the Thopaz off and on again.
What to do if Medela ThopEasy+ Medical Equipment USB connection is not permitted?
If the Medela Medical Equipment displays 'USB connection not permitted', unplug the USB cable.
What to do if Medela Medical Equipment overheated?
If the Medela Medical Equipment overheats, replace the Thopaz device.