24
Pressure sensor
Indicates whether the pressure sensor in the hydraulic circuit switches on when the
stretcher is loaded. The sensor is even activated by the weight of the stretcher itself. It
is deactivated when the stretcher rests on the arm of the loading device.
Lighten sensor
This sensor is located in the front of the stretcher. It is switched on when it is pressed
or when the stretcher rests on the arm of the loading device. When both the lighten
sensor and the locking sensor are switched on, the “+” button is deactivated.
Locking sensor
This sensor is located in the front of the stretcher. It indicates that the stretcher is
locked with the loading device.
Charging sensor
Indicates active battery charging.
Graphical representation of button activation
Used to verify the correct functioning of the “+” and “−” control buttons on the control
panel.