After the verification of the system
Perform a second calibration of the pneumotachograph and check the results.
Deviation less than 5%. Calibration is now good and the system can be used.
Deviation between 5 and 10%. The calibration is acceptable, and the system
can be used. The deviation could be a sign that device maintenance is
necessary. Please contact your Medisoft representative.
Deviation greater than 10%. The calibration is not acceptable. A verification of
the system by a field service technician is required. Please contact your
Medisoft representative.