ADVANTAGE
®
PLUS Annual Preventive
Maintenance Instructions
50097-003 Rev. A Page 1 of 12
These instructions should be followed to perform preventive maintenance on the Advantage Plus
reprocessor.
Before commencing work on the reprocessor, ensure that you have the appropriate parts and tools.
Parts Required
1. 78400-346 Advantage Plus Annual PM Kit
Tools Required
1. Philips screwdriver
2. Straight blade screwdriver
3. 2.5mm Allen wrench
4. Needle nose pliers
5. Tubing cutter
6. Pliers
7. 7mm nut driver
8. 8mm Allen wrench
9. Wire clippers
10. Adjustable wrench
11. Silicone oil
12. Hand pump
Procedure
Place the “Machine Disinfection” blocks in each basin.
Remove the alcohol and detergent containers. Purge the alcohol and detergent lines by activating the
valves/pumps Y11, Y10, P0Z, and P0C in expert mode. When the lines are purged press the “release
all” button in the expert screen. Perform on both basins.
Shutdown the computer, unplug the power to the unit and turn off the water.
Remove the access panels on the left and right sides.
Note: While performing the PM check for internal and external fluid leaks.

Other manuals for Medivators ADVANTAGE PLUS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Medivators ADVANTAGE PLUS and is the answer not in the manual?

Medivators ADVANTAGE PLUS Specifications

General IconGeneral
ModelADVANTAGE PLUS
Automated Leak TestingYes
Disinfection MethodHigh-Level Disinfection
Detergent CompatibilityEnzymatic detergents
Disinfectant CompatibilityPeracetic acid
Electrical Requirements120 VAC, 60 Hz, 15A or 230 VAC, 50 Hz, 10A
Power Requirements120 VAC, 60 Hz, 15A or 230 VAC, 50 Hz, 10A
CompatibilityFlexible Endoscopes

Related product manuals