6. Error messages/troubleshooting
If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(example: error no. 1)
Error No. Possible cause(s)/solutions
ERR 1 The systolic pressure was determined. The tube may have loos-
ened. Or no pulse was detected. * Ensure cuff connections are
tight with proper cuff placement. See instruction booklet. (5.3)
ERR 2 The pressure dropped below 20 mmHg without a measurement
taking place. The Release valve was pressed during the measure-
ment. Re-start.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated
or the hose connection is not tight. Check connections and repeat.
ERR 5 The difference between systolic and diastolic is excessive. Mea-
sure again, carefully following proper cuff procedures and ensure
measurement is taken under quiet conditions.
Other possible errors and solutions
If problems occur when using the device, the following points should be checked:
Situation Solution
The display remains empty when the instrument is Check whether the batteries are installed with
switched on although the batteries are in place. the correct polarity.
The display is not working properly. •.If the display is unusual, remove the batteries
and then exchange them for new.
The pressure does not rise although you are Check the connection of the cuff tube and
pumping the bulb. connect properly.
The device frequently fails to measure the blood 1. Fit the cuff correctly on the arm. (Section 5.3)
pressure values, or the values measured are too 2. Before starting measurement, make sure
low or high. the cuff is not too tight and clothing is not exert-
ing pressure on the arm.
Take articles of clothing off if necessary.
3. Measure again in complete peace and quiet.
Every measurement results in different values
Blood pressure changes constantly. The observed
although the device functions normally and normal readings may accurately reflect your pressure.
values are displayed. Please read the information listed under common
sources of error. (5.2) Repeat the measurement.
Blood pressure values differ from
Pressure readings in your doctors office may be
those measured by your doctor. higher due to anxiety.
Record the daily values and consult your doctor.